Transmitter M300 61
© 12 / 11 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M300
Printed in Switzerland 52 121 389
imits and is the default setting. Bi-Linear will also prompt for a scaling value for the mid-point
of the signal and allows two different linear segments between the minimum and maximum
scaling limits.
Enter the minimum and maximum Value of Aout.
If Auto-range was selected then Aout max1 can be configured. Aout max1 is the maximum
value for the first range on Auto-Range. The maximum value for the second range on Auto-
Range was set in the previous menu. If Logarithmic Range was selected, it will also prompt
for the number of decades as ”Aout1 # of Decades =2”.
The value for the Hold mode can be configured to hold the Last value or can be set to a Fixed
value.
Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will
discard the entered values and return to the measurement display screen, selecting Yes will
save changes made.