11.2.1 Calibrate meter (only for channel A)
The M400 transmitter is factory calibrated within specifications. It is not normally necessary to
perform meter re-calibration unless extreme conditions cause an out of spec operation shown by
Calibration Verification. Periodic verification / re-calibration may also be necessary to meet Q.A. re-
quirements. Meter calibration can be selected as current (used for most dissolved oxygen, Volt-
age, Rg Diagnostic, Rr Diagnostic (used for pH), and temperature (used for all measurements).
11.2.1.1 Temperature
Temperature is performed as a three point calibration. The table above shows the resistance val-
ues of these three points.
Navigate to the Calibrate Meter screen and choose Temperature calibration for Channel A.
Press [ENTER] to begin temperature calibration process
The first text line will ask for the Point 1 temperature resistance value (this will correspond to
temperature 1 value shown on the calibration module accessory). The second text line will show
the measured resistance value. When the value stabilizes, press [ENTER] to perform calibration.
The transmitter screen will then prompt the user to enter the value for Point 2, and T2 will dis-
play the measured resistance value. When this value stabilizes, press [ENTER] to calibrate this
range.
Repeat these steps for Point 3.
Press [ENTER] to bring up a confirmation screen. Select Yes to save the calibration values and
the successful calibration is confirmed on the display.
The transmitter will return to the measurement mode in approximately 5 seconds.
11.2.1.2 Current
Current calibration is preformed as a two point calibration.
Navigate to the Calibrate Meter screen and select Channel A.