Mettler Toledo M400 Specifikace Strana 35

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 88
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 34
Transmitter Multiparameter M400 35
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400
Printed in Switzerland 52 121 378
Process calibration at oxygen values below 5% air saturation is only available for M400 Type 3
in combination with InPro 6970 i.
The InPro 6880 i and InPro 6870 i only provide process calibration at oxygen values above 5%
air saturation.
7.4 pH Calibration
For pH sensors, the M400 transmitter features one-point, two-point (Auto or Manual mode)
or process calibration with 8 preset buffer sets or manual buffer entry. Buffer values refer to
25 °C. To calibrate the instrument with automatic buffer recognition, you need a standard pH
buffer solution that matches one of these values. (See section 8.2.3.2 for configuring modes
and selecting buffer sets.) Please select the correct buffer table before using automatic
calibration (see chapter 19).
After every successful calibration, the three following options are available:
Adjust: Calibration values are taken over and used for the measurement. Additionally,
the data’s will be stored in the calibration history.
Calibrate: Calibration values will be stored in the calibration history, but will be not used for
the measurement. The calibration values from the last valid Adjustment will be further
used for the measurement.
Abort: Calibration values will be aborted.
Enter pH Calibration mode as described in section 7.1.
7.4.1 One point calibration
Select 1 point Calibration.
Place the electrode in the buffer solution and press the [ENTER] key to start the calibration.
Auto mode: The display shows the buffer that the transmitter has recognized (Point 1) and the
measured value.
Manual mode: Enter the buffer value and press [ENTER] to proceed.
As soon as the drift conditions have stabilized (or [ENTER] pressed in manual mode) the
display changes to show the slope calibration factor S and the offset calibration factor Z.
After a successful calibration, the calibration values are taken over and stored in the cal history
(Adjust), only stored in the cal history (Calibrate) or aborted.
Zobrazit stránku 34
1 2 ... 30 31 32 33 34 35 36 37 38 39 40 ... 87 88

Komentáře k této Příručce

Žádné komentáře