Transmitter M400 99
© 02 / 2010 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter M400
Printed in Switzerland 52 121 378
11.1.6 Set Relay
The Set Relays diagnostic menu allows to open or close each relay manually. To access relays
5 and 6, press [ENTER].
0 = open the relay
1 = close the relay
Press [ENTER] to return to Measurement mode.
11.1.7 Read relays
The Read Relays diagnostic menu shows the state of each relay as defined below.
To display relays 5 and 6, press [ENTER]. Press [ENTER] again to exit from this display.
0 = Normal
1 = Inverted.
11.1.8 Set analog outputs
This menu enables the user to set all analog outputs to any mA value within the 0–22 mA
range. Press [ENTER] to exit from this display.
11.1.9 Read analog outputs
This menu shows the mA value of the analog outputs.