Mettler Toledo M300 FLOW Specifikace Strana 54

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 64
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 53
Transmitter O
2
M300 54
© 05/07 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter O
2
M300
Printed in Switzerland 52 121 305
12 Info
(PATH: Info)
Pressing the key will display the Info Menu with the options Messages and Calibration Data
12.1 Messages
(PATH: Info/Messages)
The most recent message is displayed. The up and down arrow keys allow scrolling through
the last four messages that have occurred.
Clear Messages clears all the messages. Messages are added to the message list when the
condition that generates the message first occurs. If all messages are cleared and a message
condition still exists and started before the clear then it will not appear in the list. For this
message to re-occur in the list the condition must go away and then reappear.
12.2 Calibration data
(PATH: Info/Calibration Data)
Selecting Calibration Data displays the slope (S) and offset calibration constants (Z).
P = calibration constants for the primary measurement
S = calibration constants for the primary measurement
Zobrazit stránku 53
1 2 ... 49 50 51 52 53 54 55 56 57 58 59 ... 63 64

Komentáře k této Příručce

Žádné komentáře